Page 24 AF910 Manual Version 1.4
Figure 27
Example (using picture above). Using this example, the home or zero position is above
focus, so raising objectives above sample will decrease motor counts, while lowering
objectives closer to sample will increase counts.
- Objective #1 is selected as the reference objective. When in focus its
motor step count position is 25,000.
- The remaining objectives have focus positions (motor steps) as follows:
o Objective #2 = 34,000.
o Objective #3 = 36,500.
o Objective #4 = 15,400.
o Objective #5 = 20,125.
o Objective #6 = 42,000.
- Each objective offset can be calculated like this:
o Objective #1 offset is zero since it is the reference objective. Any
objective number can be selected as the reference.
o Objective #2 offset: 34,000 25,000 = 9000 (or hex h2328).
o Objective #3 offset: 36,500 25,000 = 11,500 (h2CEC).
o Objective #4 offset: 15,400 25,000 = -9600 (negative, use 2’s
compliment, so hDA80).
o Objective #5 offset: 20,125 25,000 = -4875 (hECF5).
o Objective #6 offset: 42,000 25,000 = 17,000 (h4268).